Proline Cubemass C 300
Mounting
Hauser
21
6
Mounting
6.1
Mounting requirements
6.1.1
Mounting position
Mounting location
A0028772
To prevent measuring errors arising from accumulation of gas bubbles in the measuring
pipe, avoid the following mounting locations in the piping:
• Highest point of a pipeline.
• Directly upstream of a free pipe outlet in a down pipe.
Installation in down pipes
However, the following installation suggestion allows for installation in an open vertical
pipeline. Pipe restrictions or the use of an orifice with a smaller cross-section than the
nominal diameter prevent the sensor running empty while measurement is in progress.
1
2
3
4
5
A0028773
4
Installation in a down pipe (e.g. for batching applications)
1
Supply tank
2
Sensor
3
Orifice plate, pipe restriction
4
Valve
5
Batching tank