Proline Cubemass C 300
Technical data
Hauser
205
Data logging (line recorder):
• Memory capacity for up to 1000 measured values is activated.
• 250 measured values can be output via each of the 4 memory channels. The recording
interval can be defined and configured by the user.
• Measured value logs can be accessed via the local display or operating tool e.g. FieldCare,
DeviceCare or Web server.
For detailed information, see the Operating Instructions for the device.
Heartbeat Technology
Order code for "Application package", option EB "Heartbeat Verifi Monitoring"
Heartbeat Verification
Meets the requirement for traceable verification to DIN ISO 9001:2008 Chapter 7.6 a)
"Control of monitoring and measuring equipment".
• Functional testing in the installed state without interrupting the process.
• Traceable verification results on request, including a report.
• Simple testing process via local operation or other operating interfaces.
• Clear measuring point assessment (pass/fail) with high test coverage within the
framework of manufacturer specifications.
• Extension of calibration intervals according to operator's risk assessment.
Heartbeat Monitoring
Continuously supplies data, which are characteristic of the measuring principle, to an
external condition monitoring system for the purpose of preventive maintenance or
process analysis. These data enable the operator to:
• Draw conclusions - using these data and other information - about the impact the
process influences (e.g. corrosion, abrasion, formation of buildup etc.) have on
measuring performance over time.
• Schedule servicing in time.
• Monitor the process or product quality, e.g. gas pockets.
For detailed information, see the Special Documentation for the device.
Concentration
measurement
Order code for "Application package", option ED "Concentration"
Calculation and outputting of fluid concentrations.
The measured density is converted to the concentration of a substance of a binary mixture
using the "Concentration" application package:
• Choice of predefined fluids (e.g. various sugar solutions, acids, alkalis, salts, ethanol
etc.).
• Common or user-defined units (°Brix, °Plato, % mass, % volume, mol/l etc.) for standard
applications.
• Concentration calculation from user-defined tables.
For detailed information, see the Special Documentation for the device.
Special density
Order code for "Application package", option EE "Special density"
Many applications use density as a key measured value for monitoring quality or
controlling processes. The device measures the density of the fluid as standard and makes
this value available to the control system.
The "Special Density" application package offers high-precision density measurement over a
wide density and temperature range particularly for applications subject to varying process
conditions.
For detailed information, see the Operating Instructions for the device.