Proline Cubemass C 300
Commissioning
Hauser
107
10.4.13 Configuring partially filled pipe detection
The
Partial filled pipe detection
wizard guides you systematically through all parameters
that have to be set for configuring the monitoring of the pipe filling.
Navigation
"Setup" menu → Partially filled pipe detection
‣
Partially filled pipe detection
Assign process variable
Low value partial filled pipe detection
High value partial filled pipe detection
Response time part. filled pipe detect.
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Assign process variable
–
Select process variable for
partially filled pipe detection.
• Off
• Density
• Calculated
reference density
Off
Low value partial filled pipe
detection
A process variable is selected
in the
Assign process variable
Enter lower limit value for
deactivating partialy filled pipe
detection.
Signed floating-point
number
Depends on country:
• 200 kg/m
3
• 12.5 lb/ft
3
High value partial filled pipe
detection
A process variable is selected
in the
Assign process variable
Enter upper limit value for
deactivating partialy filled pipe
detection.
Signed floating-point
number
Depends on country:
• 6
000 kg/m
3
• 374.6 lb/ft
3
Response time part. filled pipe
detect.
A process variable is selected
in the
Assign process variable
Use this function to enter the
minimum time (hold time) the
signal must be present before
diagnostic message S962 "Pipe
only partly filled" is triggered in
the event of a partially filled or
empty measuring pipe.
0 to 100 s
1 s
Summary of Contents for Proline Cubemass C 300