Proline Cubemass C 300
Technical data
Hauser
201
Supported functions
Data exchange between the operating unit (such as a notebook for example) and the
measuring device:
• Upload the configuration from the measuring device (XML format, configuration
backup)
• Save the configuration to the measuring device (XML format, restore configuration)
• Export event list (.csv file)
• Export parameter settings (.csv file or PDF file, document the measuring point
configuration)
• Export the Heartbeat verification log (PDF file, only available with the "Heartbeat
Verification" application package)
• Flash firmware version for device firmware upgrade, for instance
• Download driver for system integration
• Visualize up to 1000 saved measured values (only available with the
Extended
HistoROM
Web server special documentation → 207
HistoROM
data management
The measuring device features HistoROM data management. HistoROM data management
comprises both the storage and import/export of key device and process data, making
operation and servicing far more reliable, secure and efficient.
When the device is delivered, the factory settings of the configuration data are stored
as a backup in the device memory. This memory can be overwritten with an updated
data record, for example after commissioning.
Additional information on the data storage concept
There are different types of data storage units in which device data are stored and used by the device:
HistoROM backup
T-DAT
S-DAT
Available data
• Event logbook such as diagnostic
events for example
• Parameter data record backup
• Device firmware package
• Measured value logging ("Extended
HistoROM" order option)
• Current parameter data record (used by
firmware at run time)
• Maximum indicators (min/max values)
• Totalizer values
• Sensor data: nominal diameter etc.
• Serial number
• Calibration data
• Device configuration (e.g. SW options,
fixed I/O or multi I/O)
Storage location
Fixed on the user interface board in
the connection compartment
Attachable to the user interface board in the
connection compartment
In the sensor plug in the transmitter neck
part
Data backup
Automatic
• The most important device data (sensor and transmitter) are automatically saved in the
DAT modules
• If the transmitter or measuring device is replaced: once the T-DAT containing the
previous device data has been exchanged, the new measuring device is ready for
operation again immediately without any errors
• If the sensor is replaced: once the sensor has been replaced, new sensor data are
transferred from the S-DAT in the measuring device and the measuring device is ready
for operation again immediately without any errors
• If exchanging the electronics module (e.g. I/O electronics module): Once the electronics
module has been replaced, the software of the module is compared against the current
device firmware. The module software is upgraded or downgraded where necessary. The
electronics module is available for use immediately afterwards and no compatibility
problems occur.