Proline Cubemass C 300
Operation
Hauser
147
Parameter overview with brief description
Parameter
Prerequisite
Description
User interface / User
entry
Factory setting
Inhomogeneous medium index
–
Shows the degree of
inhomogeneity of the medium.
Signed floating-point
number
–
Cut off inhomogeneous wet gas
–
Enter cut off value for wet gas
applications. Below this value
the 'Inhomogeneous medium
index' is set to 0.
Positive floating-
point number
0.25
Cut off inhomogeneous liquid
–
Enter cut off value for liquid
applications. Below this value
the 'Inhomogeneous medium
index' is set to 0.
Positive floating-
point number
0.05
Suspended bubbles index
The diagnostic index is only
available for Promass Q.
Shows the relative amount of
suspended bubbles in the
medium.
Signed floating-point
number
–
Cut off suspended bubbles
The parameter is only available
for Promass Q.
Enter the cut off value for
suspended bubbles. Below this
value the 'Index for suspended
bubbles' is set to 0.
Positive floating-
point number
0.05