Diagnostics and troubleshooting
Proline Cubemass C 100 HART
94
Hauser
Parameter
Description
User interface
Factory setting
IP address
Displays the IP address of the Web server of
the measuring device.
4 octet: 0 to 255 (in the
particular octet)
192.168.1.212
Subnet mask
Displays the subnet mask.
4 octet: 0 to 255 (in the
particular octet)
255.255.255.0
Default gateway
Displays the default gateway.
4 octet: 0 to 255 (in the
particular octet)
0.0.0.0
12.11 Firmware history
Release
date
Firmware
version
Order code
for
"Firmware
version"
Firmware
changes
Documentation
type
Documentation
04.2013 01.00.00
Option
76
Original firmware
Operating
Instructions
BA01188D/06/EN/01.13
06.2014 01.01.zz
Option
70
• In accordance with
HART 7
Specification
• Integration of
optional onsite
display
• New unit "Beer
Barrel (BBL)"
• Monitoring of
measuring tube
damping
• Simulation of
diagnostic events
• External
verification of the
current and PFS
output via the
Heartbeat
application package
• Fixed value for
simulation pulses
Operating
Instructions
BA01188D/06/EN/02.14
Flashing the firmware to the current version or to the previous version is possible via
the service interface (CDI) .
For the compatibility of the firmware version with the previous version, the installed
device description files and operating tools, observe the information about the device
in the "Manufacturer's information" document.
The manufacturer's information is available:
• In the Download Area of the Hauser Internet site: www.endress.com →
Download
• Specify the following details:
• Product root, e.g. 8E1B
• Text search: Manufacturer's information
• Search range: documentation
Summary of Contents for Proline Cubemass C 100 HART
Page 126: ...www addresses endress com...