Proline Cubemass C 100 HART
Technical data
Hauser
101
16.3 Input
Measured variable
Direct measured variables
• Mass flow
• Density
• Temperature
Calculated measured variables
• Volume flow
• Corrected volume flow
• Reference density
Measuring range
Measuring ranges for liquids
DN
Measuring range full scale values
min(F)
to
max(F)
[mm]
[in]
[kg/h]
[lb/min]
1
¹⁄₂₄
0 to 20
0 to 0.735
2
¹⁄₁₂
0 to 100
0 to 3.675
4
¹⁄₈
0 to 450
0 to 16.54
6
¹⁄₄
0 to 1
000
0 to 36.75
Measuring ranges for gases
The full scale values depend on the density of the gas and can be calculated with the
formula below:
max(G)
=
max(F)
· ρ
G
: x
max(G)
Maximum full scale value for gas [kg/h]
max(F)
Maximum full scale value for liquid [kg/h]
max(G)
<
max(F)
max(G)
can never be greater than
max(F)
ρ
G
Gas density in [kg/m³] at operating conditions
x
Constant dependent on nominal diameter
DN
x
[mm]
[in]
[kg/m
3
]
1
¹⁄₂₄
20
2
¹⁄₁₂
20
4
¹⁄₈
20
6
¹⁄₄
20
Recommended measuring range
Operable flow range
Over 1000 : 1.
Flow rates above the preset full scale value do not override the electronics unit, with the
result that the totalizer values are registered correctly.