Micropilot S FMR540 4 to 20 mA HART
Technical data
Hauser
71
10.1.7
Operating conditions: Process
Process temperature range
FKM Viton GLT, -40 °C to +200 °C (-40 °F to +392 °F)
Dielectric constant
In free space:
ε
r
≥
1.8
10.1.8
Mechanical construction
Weight
Approx 6 kg + weight of flange
Material
Refer to TI00412F/00/EN, chapter "Material (not in contact with process)" and "Material (in
contact with process).
10.1.9
Certificates and approvals
CE approval
The measuring system meets the legal requirements of the EC-guidelines. Hauser confirms
the device passing the required tests by attaching the CE-mark.
RF approvals
R&TTE, FCC
External standards and
guidelines
EN 60529
Protection class of housing (IP-code).
EN 61010
Safety regulations for electrical devices for measurement, control, regulation and laboratory use.
EN 61326
Emissions (equipment class B), compatibility (appendix A - industrial area).
NAMUR
Standards committee for measurement and control in the chemical industry.
Ex approval
Correlation of safety instructions (XA) and certificates (ZE) to the device:
Control Drawings
Correlation of Control Drawings to the device:
10.1.10 Supplementary Documentation
Supplementary
Documentation
• Technical Information (TI00412F/00/EN)
• Operating Instructions "Description of Instrument functions" (BA00341F/00/EN)
• Brief Operating Instructions (KA01059F/00/EN)
Device
Certificate
Explosion protection
Output
Communication
PTB 00 ATEX
XA
WHG
FMR540
1
ATEX II 1/2 G EEx ia IIC T6
A
HART
2067X
XA00338F/00/A3
ZE00243F/00/DE
6
ATEX II 1/2 G EEx ia, WHG
Device
Certificate
Explosion protection
Output
Communication
Documentation
FMR540
S
FM IS
A
HART
XA00554F/00/EN
U
CSA IS
ZD00196F/00/EN