Micropilot S FMR540 4 to 20 mA HART
Installation
Hauser
21
L00-FMR540xx-00-00-00-yy-010
3.
Place the Sensor alignment tool (a) for Micropilot S
FMR540. Please, note to avoid any obstacles between
the backside of the Alignment tool and the nameplate
of Micropilot S FMR540.
L00-FMR540xx-00-00-00-yy-007
4.
Micropilot S FMR540 with Horn Antenna:
Tilt the FMR540 targeting the direction of tank center
up to the position where the angle indicators`s outer
circle reaches the circle of 3 deg (e).
Note!
Exceeding the 3 degree position may cause a weaker
signal (or loss of signal).
Micropilot S FMR540 with Parabolic Antenna:
Tilt the FMR540 to the position where the bubble
comes into the center (f) of the inclination indicator (0
deg).
Gradually tighten the nut (b) of the alignment tool and
make sure to keep the position of 0 degree/3 degree
inclination.
After tightening the nut, check if the sensor cannot tilt
and change its position.
Torque for the nut: 80 to 85 Nm
If it is required by the local custody transfer authority,
please seal the alignment device at the sealing screws
(c) using the provided wires and seal metals.
a
90°
90°
e
f
3°
2
°
1°
0
°
E
ndress
+H
auser
52026756