Micropilot S FMR540 4 to 20 mA HART
Installation
Hauser
17
Standard installation FMR540 with horn
antenna
• Observe installation instructions,
• Marker must be aligned towards tank wall.
The marker is located clearly visible on the
sensor neck or the flange.
• After mounting, the housing can be turned
350° in order to simplify access to the display
and the terminal compartment.
• Adjust vertical sensor alignment in case the
flange is not parallel to the face is medium
surface.
• The horn antenna should protrude from the
nozzle. If necessary, choose version with
antenna extension (
11).
Note!
Please contact Hauser for application
with higher nozzle.
• The horn antenna should be installed with 3°
inclination towards the tank center.
To avoid interference reflections or for
optimum alignment within the tank, the
FMR540 with optional alignment device can
be swiveled by 15° in all directions. For more
informations please see Operating Instructions
KA00274F/00/A2.
Please contact Hauser Service
Organization for commissioning.
L00-FMR540xx-17-00-00-xx-001
H
Ø
D
Antenna size
100 mm (4")
D [mm (in)]
95 (3.7)
H [mm (in)]
(without antenna extension)
< 430 (< 19.2)