Electrical connection
Liquistation CSF48
44
Hauser
6.4
Terminal assignment for input/output signals
The following signals can be configured for external sampler control:
Input signals
• 2 analog signals 0/4 to 20 mA
• 2 binary signals > 100 ms pulse width or edge
• Signals of digital sensors with Memosens protocol (optional)
Output signals
• 2 Binary signals > 1 s pulse width or edge
The controller must be opened to allow the signal cable, sensor cable and optional relay to be
connected.
6.5
Connecting the sensors
6.5.1
Sensor types with Memosens protocol
a0016348
Fig. 51: Connection example for binary output with relay
Sensor types
Sensor cable
Sensors
Digital sensors
without
additional
internal power supply
CYK10 with plug connection
and inductive signal
transmission
• pH sensors
• ORP sensors
• Combined sensors
• Amperometric oxygen sensors
• Conductive conductivity sensors
• Chlorine sensors
Fixed cable
Inductive conductivity sensors
Digital sensors
with
additional
internal power supply
Fixed cable
• Turbidity sensors
• Sensors for interface measurement
• Sensors for measuring the spectral absorption
coefficient (SAC)
• Nitrate sensors
• Optical oxygen sensors
• Ion-sensitive sensors