
Operation
Liquistation CSF48
104
Hauser
Settings with the Basic program type with multiple bottles
Sampling mode = "Time paced CTCV"
Assignment bin. output
Options
• No binary output
configured
• Binary output S:x
Factory setting
No binary output configured
Assignment of the binary output to the program cycle.
Inputs
Settings for the inputs can be made as described in the
"Inputs" section.
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function
Options
Info
Sampling interval (for version
with vacuum pump or
peristaltic pump)
Sampling interval (for version
with sampling assembly)
00:01:00 to 99:59:00
HH:MM:SS
Factory setting
00:10:00 HH:MM:SS
00:00:10 to 99:59:00
HH:MM:SS
Factory setting
00:10:00 HH:MM:SS
Set the sampling interval.
Dosing volume (for version
with vacuum pump or
sampling assembly)
Sampling volume (for version
with peristaltic pump)
Vacuum pump:
20 to 350 ml
Peristaltic pump:
10 to 10000 ml
Sampling assembly:
10 to 1000 ml
Factory setting
Vacuum pump:
200 ml
Peristaltic pump:
100 ml
Sampling assembly:
200 ml
Set the dosing volume or the sampling volume.
The volume is taken from the setup in the version
with a vacuum pump or sampling assembly.
The dosing accuracy and the repeatability of a sample
volume < 20 ml can vary, depending on the specific
application.
Multiplier
(only for version with vacuum
pump or sampling assembly)
1 to 10
Factory setting
1
You can change the sampling volume using the
multiplier.
For example, if the dosing volume is set to 200 ml, set
the multiplier to 2 for a sampling volume of 400 ml. 2
samples are taken in succession when sampling.
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function
Options
Info
Summary of Contents for Liquistation CSF48
Page 2: ......
Page 267: ...Liquistation CSF48 Endress Hauser ...
Page 268: ...www addresses endress com ...