Liquistation CSF48
Operation
Hauser
181
Conc. Table
Options
• NaOH 0..15%
• HCl 0..20%
• HNO3 0..25%
• HNO3 24..30%
• H2SO4 0..28%
• H2SO4 40..80%
• H2SO4 93..100%
• H3PO4 0..40%
• NaCl 0..26%
• User table 1
• User table 2
• User table 3
• User table 4
Factory setting
NaOH 0..15%
Concentration tables saved at the factory:
NaOH: 0 to 15%, 0 to 100
˚
C (32 to 212
˚
F)
HCl: 0 to 20%, 0 to 65
˚
C (32 to 149
˚
F)
HNO
3
: 0 to 25%, 2 to 80
˚
C (36 to 176
˚
F)
H
2
SO
4
: 0 to 28%, 0 to 100
˚
C (32 to 212
˚
F)
H
2
SO
4
: 40 to 80%, 0 to 100
˚
C (32 to 212
˚
F)
H
2
SO
4
: 93 to 100%, 0 to 100
˚
C (32 to 212
˚
F)
H
3
PO
4
: 0 to 40%, 2 to 80
˚
C (36 to 176
˚
F)
NaCl: 0 to 26%, 2 to 80
˚
C (36 to 176
˚
F)
Compensation
Options
• None
• Linear
• NaCl (IEC 746-3)
• Water ISO7888 (20°C)
• Water ISO7888 (25°C)
• UPW NaCl
• UPW HCl
• User table 1
• User table 2
• User table 3
• User table 4
Factory setting
Linear
Various methods are available to compensate for the
temperature dependency.
Depending on your process, decide which type of
compensation you want to use.
Alternatively, you can also select "None" and thus
measure uncompensated conductivity.
Current output
Range lower value
0 to 100 %
Operating mode = "Concentration"
Range upper value
Range lower unit
Options
• S/m
•
μ
S/cm
• mS/cm
• S/cm
•
μ
S/m
• mS/m
Factory setting
mS/cm
Operating mode = "Conductivity"
Define the units and range limits for the output values.
Range lower value
Enter value
Range upper unit
Enter value
Range upper value
Enter value
Path: Menu/Setup/Additional functions/Measuring range switch
Function
Options
Info