Liquistation CSF48
Operation
Hauser
117
Settings with a Basic program and external signal
Settings with the Basic program type with 1 bottle
Sampling mode = "External signal"
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function
Options
Info
Sampling volume
10 to 1000 ml
Factory setting
100 ml
Enter the sample volume.
Sampling signal input
Options
• No sampling input
configured
Factory setting
No sampling input configured
Select the input for the sampling signal.
The fieldbus must be configured for this function.
The sampling input can be configured under Inputs
.
Bottle change mode
Options
• Number of samples
• Time
• External signal
Factory setting
Number of samples
The bottle can be changed after a specific number of
samples, after a time or by an external signal.
If Bottle change mode
Number of samples
is selected:
Samples per bottle
1 to 9999
Factory setting
1
Set the number of samples.
If Bottle change mode
Time
is selected:
Time interval
00-00:02 to 31-00:00
DD-HH:MM
Factory setting
00-01:00 DD-HH:MM
Set the time (days, hours and minutes) after which
the system should change to the next bottle.
Bottle synchronization
Options
• None
• 1. bottle change time
• 1. Time of bottle
number
Factory setting
None
None:
The time of sampling and bottle change are not
synchronized.
1. bottle change time:
Sampling starts with the first bottle.
Set the synchronization time.
1. Time of bottle number:
Each bottle is assigned to a specific fill time.
Set the synchronization time and the weekday.