Liquistation CSF48
Diagnostics and troubleshooting
Hauser
209
11.11 Simulation
You can simulate values at inputs and outputs for testing purposes:
• Current values at current outputs
• Measured values at inputs
• Relay contact opening or closing
Only current values are simulated. It is not possible to use the simulation function to
calculate the totalized value for the flow or rainfall.
The inputs and outputs must be activated in the Setup menu prior to simulation.
Diagnostics/Simulation
Function
Options
Info
Current output x:y
Simulation of an output current
This menu appears once for each current output.
Simulation
Options
• On
• Off
Factory setting
Off
If you simulate the value at the current output, this is
indicated on the display by a simulation icon in front of
the current value.
Current
2.4 to 23.0 mA
Factory setting
4 mA
Set the desired simulation value.
Alarm relay
Relay x:y
Simulation of a relay state
This menu appears once for each relay.
Simulation
Options
• On
• Off
Factory setting
Off
If you simulate the relay state, this is indicated on the
display by a simulation icon in front of the relay display.
State
Options
• Low
• High
Factory setting
Low
Set the desired state.
The relay switches in accordance with your setting when
you switch on the simulation. The display shows "On"
(="Low") or "Off" (="High") for the simulated relay state.
Meas. inputs
Simulation of a measured value
This menu appears once for each measuring input.
Channel : parameter
Sim. main value
Options
• On
• Off
Factory setting
Off
If you simulate the measured value, this is indicated on
the display by a simulation icon in front of the measured
value.