Liquistation CSF48
Diagnostics and troubleshooting
Hauser
199
11.9 Logbooks
11.9.1 Available logbooks
Types of logbooks
• Logbooks physically available (all apart from the overall logbook)
• Database view of all logbooks (=overall logbook)
11.9.2 Logbooks menu
Logbook
Visible in
Max.
entries
Can be
disabled
1)
1)
Data in brackets means this depends on the overall logbook
Logbook
can be
deleted
Entries
can be
deleted
Can be
exported
Overall logbook
All events
1000
Yes
No
Yes
No
Calibration logbook
Calibration events
75
(Yes)
No
Yes
Yes
Operation logbook
Configuration events
250
(Yes)
No
Yes
Yes
Diagnostics logbook
Diagnostic events
250
(Yes)
No
Yes
Yes
Program logbook
Program logbook
5000
Yes
No
Yes
Yes
Version logbook
All events
50
No
No
No
Yes
Hardware version
logbook
All events
125
No
No
No
Yes
Data logbook for sensors
(optional)
Data logbooks
150,000
Yes
Yes
Yes
Yes
Debugging logbook
Only accessible with the
special activation code
(Service)
1000
Yes
No
Yes
Yes
Diagnostics/Logbooks
Function
Options
Info
Program logbook
Chronological list of the programming events.
Show
Events are displayed
Select a particular event to display more detailed
information.
Go to date
User input
• Go to date
• Time
Use this function to go directly to a specific time in the
list. In this way, you avoid having to scroll through all
the information. The complete list is always visible,
however.
Show summary of
current program
Read only
The bottle statistics for the sampler are displayed. The
statistics appear for each individual bottle after the
program starts.
Further information is provided in the "Bottle statistics"
section.