Electrical connection
Liquiline CM442R/CM444R/CM448R
36
Hauser
PWR
Power
GN
Supply voltage is applied and module is initialized
BF
Bus failure
RD
Not used
SF
System failure
RD
System failure
COM
Communication
YE
Sending or receiving Modbus message
DIP switches on front of module
DIP
Factory setting
Assignment
1-128
ON
Not used
OFF
Write protection: "ON" = configuration
not
possible via Ethernet, only via local operation
Service OFF
When the switch is set to "
ON"
, the user settings for Ethernet addressing are saved and
connection settings programmed at the factory are activated:
IP address=192.168.1.212, subnet mask=255.255.255.0, gateway=0.0.0.0, DHCP=off
When the switch is set to "
OFF"
, the saved user settings are active again.
LEDs on front of module
LED
Designation
Color
Description