Technical data
Liquiline CM442R/CM444R/CM448R
154
Hauser
15.6.4 Cable specification
Cable type
Recommended: shielded cable
Cross-section
Max. 2.5 mm
2
(14 AWG)
15.7 Relay outputs
15.7.1 Relay types
• 1 single-pin changeover contact (alarm relay)
• 2 or 4 one-pin changeover contacts, (optionally with extension modules)
15.7.2 Relay switching capacity
Base module (alarm relay)
Switching voltage
Load (max.)
Switching cycles (min.)
230 V AC, cos
= 0.8 to 1
0.1 A
700,000
0.5 A
450,000
115 V AC, cos
= 0.8 to 1
0.1 A
1,000,000
0.5 A
650,000
24 V DC, L/R = 0 to 1 ms
0.1 A
500,000
0.5 A
350,000
Extension modules
Switching voltage
Load (max.)
Switching cycles (min.)
230 V AC, cos
= 0.8 to 1
0.1 A
700,000
0.5 A
450,000
2 A
120,000
115 V AC, cos
= 0.8 to 1
0.1 A
1,000,000
0.5 A
650,000
2 A
170,000
24 V DC, L/R = 0 to 1 ms
0.1 A
500,000
0.5 A
350,000
2 A
150,000