Liquiline CM442R/CM444R/CM448R
Technical data
Hauser
155
Minimum load (typical)
• Min. 100 mA for 5 V DC
• Min. 1 mA for 24 V DC
• Min. 5 mA for 24 V AC
• Min. 1 mA for 230 V AC
15.7.3 Cable cross-section
Max. 2.5 mm
2
(14 AWG)
15.8 Protocol-specific data
15.8.1 HART
15.8.2 PROFIBUS DP
Manufacturer ID
11
h
Device type
119C
h
(CM44x), 119D
h
(CSFxx)
Device revision
001
h
HART version
7.2
Device description files (DD/DTM)
www.endress.com
Device Integration Manager (DIM)
Device variables
16 user-definable and 16 predefined device variables, dynamic variables
PV, SV, TV, QV
Supported features
PDM DD, AMS DD, DTM, Field Xpert DD
Manufacturer ID
11
h
Device type
155Dh (CM44x)
Profile version
3.02
Device master files (GSD)
www.endress.com/profibus
Device Integration Manager DIM
Output
16 AI blocks, 8 DI blocks
Input
4 AO blocks, 8 DO blocks
Supported features
• 1 MSCY0 connection (cyclic communication, master Class 1 to slave)
• 1 MSAC1 connection (acyclic communication, master Class 1 to slave)
• 2 MSAC2 connections (acyclic communication, master Class 2 to
slave)
• Device lock: the device can be locked via the hardware or software.
• Addressing using DIL switches or via the software
• GSD, PDM DD, DTM