Liquiline CM442R/CM444R/CM448R
Operation
Hauser
109
Dual conductivity
You can subtract two conductivity values and use the result, for example, to monitor the
efficiency of an ion exchanger.
Path: Menu/Setup/Additional functions/Mathematical functions/MF 1 to 6/Mode/Dual conductivity
Function
Options
Info
Calculation
Options
• Off
• On
Factory setting
Off
On/off switch for the function
Inlet
The options depend on the
sensors connected
Select the sensors that should be used as the minuend
(Inlet, e.g. sensor upstream from the ion exchanger) and
subtrahend (Outlet, e.g. sensor downstream from the
ion exchanger).
Measured value
Outlet
Measured value
Main value format
Options
• Auto
• #
• #.#
• #.##
• #.###
Factory setting
Auto
Cond. unit
Options
• Auto
•
μ
S/cm
• mS/cm
• S/cm
•
μ
S/m
• mS/m
• S/m
Factory setting
Auto
Dual conductivity
Read only
View this value in a user-defined measuring screen or
output the value via the current output.