Diagnostics and troubleshooting
Liquiline CM442R/CM444R/CM448R
126
Hauser
11.10 Simulation
You can simulate values at inputs and outputs for testing purposes:
• Current values at current outputs
• Measured values at inputs
Save logbooks
File format
Options
• CSV
• FDM
Save the logbook in the preferred file format.
You can then open the CSV file you saved on the PC in
MS Excel, for example, and process it further
1)
.
You can import the FDM files into Fieldcare and archive
them so they are tamper-proof.
All data logbooks
Data logbook 1 to n
All event logbooks
Calibration logbook
Diagnostic logbook
Configuration logbook
HW version logbook
Version logbook
The action commences as
soon as the option is
selected
Use this function to save the logbook to an SD card.
‣
Insert the SD card into the device card reader and
select the logbook to be saved.
The file name is made up of the "Logbook ident" (Menu/Setup/General settings/Logbooks), an abbreviation for
the particular logbook and a time stamp.
1)
CSV files use international number formats and separators. Therefore they must be imported into MS Excel as
external data with the correct format settings. If you double-click the file to open it, the data are only displayed
correctly if MS Excel is installed with the US country setting.
Diagnostics/Simulation
Function
Options
Info
Current outputx:y
Simulation of an output current
This menu appears once for each current output.
Simulation
Options
• On
• Off
Factory setting
Off
If you simulate the value at the current output, this is
indicated on the display by a simulation icon in front of
the current value.
Current
2.4 to 23.0 mA
Factory setting
4 mA
Set the desired simulation value.
Diagnostics/Logbooks
Function
Options
Info