Liquiline System CA80SI
Mounting
Hauser
17
54 (2.13)
49 (1.93)
15,5
(0.61)
63,5 (2.5)
A0036664
15
Dimensions of angle bracket for pressure relief valve
--- Fasteners (2 x M5)
196
(7.72)
355 (13.98)
8 (0.31)
95
(3.74)
335 (13.19)
176 (6.93)
A0036389
16
CA80SI 4-/6-channel version: Panel with pressure-reducing valves and filters, dimensions in mm (in)
2
3
0
(9
.0
6
)
255 (10.04)
8 (0.31)
74
1
5
0
(5
9
1
)
235
(9.25)
(2.91)
A0036390
17
CA80SI 4-/6-channel version: Panel with sample channel switching, dimensions in mm (in)
5.1.2
Mounting location
Note the following when erecting the device:
‣
If mounting on a wall, make sure that the wall has sufficient load-bearing capacity and
is fully perpendicular.
‣
If mounting on a base, erect the device on a level surface.
‣
Protect the device against additional heating (e.g. from a heating system).
‣
Protect the device against mechanical vibrations.
‣
Protect the device against corrosive gases, e.g. hydrogen sulfide (H
2
S) and chlorine
gases.
‣
Make sure to pay attention to the maximum height difference and the maximum
distance from the sampling point.
‣
Ensure that sample outlet hose "D" and outlet hose "W" can drain freely , without any
siphoning effects.
‣
Make sure air can circulate freely at the front of the housing.
‣
Open analyzers (i.e. analyzers that are supplied without a door) may only be erected in
closed areas or in a protective cabinet or similar facility.