Operation
Liquiline System CA80SI
68
Hauser
Menu/Setup/Analyzer/Extended setup
Function
Options
Info
Device range
Selection
The options available depend on the device
version ordered.
‣
Select the desired measuring range from the
list.
Signals
Menu/Setup/Analyzer/Extended setup
Function
Options
Info
Signal for process access
SP1
Selection
• Always accessible
•
Binary input
x:y
Factory setting
Always accessible
Always accessible
: The input signal level at the
binary inputs does not affect actions that require
sample (measurement, calibration, cleaning).
Binary input
x:y: When the signal is active at
the selected input, the analyzer starts actions
where sample is required. Otherwise the
analyzer defers or skips the action.
In the case of multi-channel devices,
SP2
to
SP6
is also displayed, depending on the number of
channels.
Diagnostic settings
Limit values for dispensers
Menu/Setup/Analyzer/Extended setup/Diagnostics settings/Limits dispensers
Function
Options
Info
Control
Selection
• Off
• On
Factory setting
On
Warning limits
Diag. code 733
Remaining operating time
Selection
1 to 90 (d)
Factory setting
28 d
A warning limit can be configured for every
dispenser:
• Dispenser 1
• Dispenser 2
• Dispenser 3
Alarm limits
Diag. code 732
Remaining operating time
Selection
1 to 60 d
Factory setting
7 d
An alarm limit can be configured for every
dispenser:
• Dispenser 1
• Dispenser 2
• Dispenser 3
Bottles
You can monitor the levels of all the bottles.
1. Switch on monitoring.
2. Enter the starting volume for every bottle.
3. Specify alarm and warning limits for the residual volumes.
4. Specify alarm and warning limits for the remaining time until the bottle refills are
consumed.