Liquiline System CA80SI
Product description
Hauser
11
3.3
Equipment architecture
3.3.1
Slot and port assignment
BASE2-E
1
2
1
2
1
2
Slot
Ports
A0044868
6
Slot and port assignment
A0044869
7
Slot and port assignment
A0040671
8
Slots and ports on the
display
*
Analyzer measured value
(parameter-specific)
• Inputs are assigned to measuring channels in the ascending order of the slots and ports.
In the example above:
"CH1: 1:1 pH glass" means:
Channel 1 (CH1) is slot 1 (basic module) : Port 1 (input 1), pH glass sensor
• Outputs and relays are named according to their function, e.g. "current output", and are
displayed in ascending order with the slot and port numbers
• Display shows SP1: analyzer measuring channel 1 with sampling point SP1 (measured
value display is parameter-specific; is not illustrated in the example)
3.3.2
Terminal diagram
The unique terminal name is derived from:
Slot no. : Port no. : Terminal
Example, NO contact of a relay
Device with 4 inputs for digital sensors, 4 current outputs and 4 relays
• Base module BASE2-E (contains 2 sensor inputs, 2 current outputs)
• 2DS module (2 sensor inputs)
• 2AO module (2 current outputs)
• 4R module (4 relays)