Electrical connection
Liquiline System CA80SI
32
Hauser
6.3.4
Connecting PROFIBUS DP or Modbus RS485
Module 485DP
A0050399
45
Module
81
82
DIP switch
on
1
T
COM
SF
BF
PWR
DGND
VP
96'
95'
99'
Termi-
nation
DP
96
95
99
2
4
8
16
32
64
128/SW
Service
A0050400
46
Wiring
diagram
Terminal
PROFIBUS DP
95
A
96
B
99
Not connected
82
DGND
81
VP
LEDs on front of module
LED
Designation
Color
Description
PWR
Power
GN
Supply voltage is applied and module is initialized.
BF
Bus failure
RD
Bus failure
SF
System failure
RD
Device error
COM
Communication
YE
PROFIBUS message sent or received.
T
Bus termination
YE
• Off = No termination
• On = Termination is used
DIP switches on front of module
DIP
Factory setting
Assignment
1-128
ON
Bus address (→ "Commissioning/communication")
OFF
Write protection: "ON" = configuration not possible via the bus, only via local
operation
Service
OFF
The switch has no function