Liquiline System CA80SI
Diagnostics and troubleshooting
Hauser
135
11.13 Device test
11.13.1 Analyzer
DIAG/System testAnalyzer/Photometer
Function
Options
Info
Cleaning factor
Read only
Raw value
Read only
DIAG/System testAnalyzer
Function
Options
Info
Cuvette heating
Cuvette heating test
Temperature
Displays the current cuvette heating temperature
Current
Displays the current heating current in ampere
Power
Displays the heating power in % (current
controller actuating value)
State
Selection
• Heating
• Heating up
• Off
Displays the cuvette heating status
Off for 10 minutes
Cuvette heating is switched off for 10 minutes.
Cuvette heating is switched on by default. It can
be switched off temporarily for testing.
On
Cuvette heating is switched on.
Magnetic stirrer
Start for 1 minute
The magnetic stirrer is switched on for 1 minute.
Stop
Sample pre-heating
Temperature
Displays the current sample pre-heating
temperature
Power
Displays the heating power in % (current
controller actuating value)
State
Displays the current sample pre-heating status
Off for 10 minutes
Sample pre-heating is switched off for 10
minutes.
On
Sample pre-heating is switched on.
Valves
Valve selection
Selection
• None
• Reagent RB
• Reagent RK
• Reagent RN
Open
Opens the selected valve.
Close
Closes the selected valve.
Pump
Start for 1 minute
The pump is switched on for 1 minute.
Summary of Contents for Liquiline CA80SI