Diagnostics and troubleshooting
Levelflex FMP55 HART
92
Hauser
3. Click "Save as ..." and save a PDF of the protocol.
The protocol contains the diagnostic messages and remedy information.
C: Via the "Eventlist / Extended HistoROM" function
1.
Select the "Eventlist / Extended HistoROM" function.
2.
Select the "Load Eventlist" function.
The list of events, including remedy information, is shown in the "Data overview"
window.
13.4 Diagnostic list
In the
Diagnostic list
submenu submenu, up to 5 currently pending diagnostic messages
can be displayed. If more than 5 messages are pending, the messages with the highest
priority are shown on the display.
Navigation path
Diagnostics → Diagnostic list
Calling up and closing the remedial measures
1. Press .
The message for the remedial measures for the selected diagnostic event opens.
2. Press + simultaneously.
The message about the remedial measures closes.