Operating menu
Levelflex FMP55 HART
134
Hauser
Additional information
DC
1
DC
1
A0013181
DC1 Relative dielectric constant of the upper medium.
For dielectric constants (DC values) of many media commonly used in various
industries refer to:
• the Hauser DC manual (CP01076F)
• the Hauser "DC Values App" (available for Android and iOS)
Interface
Navigation
Setup → Interface
Prerequisite
Operating mode (→ 128)
Interface
or
Interface with capacitance
Description
Displays the measured interface level L
I
(before linearization).
Additional information
0%
0%
L
I
L
I
A0013197
The unit is defined in the
Level unit