Operating menu
Levelflex FMP55 HART
136
Hauser
Additional information
Meaning of the options
•
Manual map
To be selected if the range of mapping is to be defined manually in the
Mapping end
point
parameter (→ 137). In this case it is not necessary to confirm the distance.
•
Distance ok
To be selected if the measured distance matches the actual distance. The device
performs a mapping.
•
Distance unknown
To be selected if the actual distance is unknown. A mapping can not be performed in this
case.
•
Distance too small
To be selected if the measured distance is smaller than the actual distance. The device
searches for the next echo and returns to the
Confirm distance
parameter. The distance
is recalculated and displayed. The comparison must be repeated until the displayed
distance matches the actual distance. After this, the recording of the map can be started
by selecting
Distance ok
.
•
Distance too big
To be selected if the measured distance exceeds the actual distance. The device adjusts
the signal evaluation and returns to the
Confirm distance
parameter. The distance is
recalculated and displayed. The comparison must be repeated until the displayed
distance matches the actual distance. After this, the recording of the map can be started
by selecting
Distance ok
.
•
Tank empty
To be selected if the tank is completely empty. The device records a mapping covering
the complete measuring range.
To be selected if the tank is completely empty. The device records a mapping covering
the complete measuring range minus
Map gap to LN
.
•
Factory map
To be selected if the present mapping curve (if one exists) is to be deleted. The device
returns to the
Confirm distance
parameter and a new mapping can be recorded.
When operating via the display module, the measured distance is displayed together
with this parameter for reference purposes.
For interface measurements the distance always refers to the toatal level (not the
interface level).
For FMP55 with rope probes and
Operating mode (→ 128) = Interface with
capacitance
the mapping must be recorded with the tank being empty, and the
Tank
empty
option must be selected. Otherwise the device can not register the correct
empty capacitance.
For FMP55 with coax probes a mapping must be recorded at least in the upper part of
the probe, as tightening the flange has an influence on the envelope curve. However,
even with coax probes it is recommended to record the mapping with the tank being
completely empty (and selecting the
Tank empty
option).
If the teaching procedure with the
Distance too small
option or the
Distance too big
option is quit before the distance has been confirmed, a map is
not
recorded and the
teaching procedure is reset after 60 s.
Present mapping
Navigation
Setup → Present mapping
Description
Indicates up to which distance a mapping has already been recorded.
7)
Only available for "Expert → Sensor → Echo tracking →
Evaluation mode
parameter" = "Short time history" or "Long time history"