Commissioning
Gammapilot M FMG60 PROFIBUS PA
66
Hauser
6.4.3
Full and empty calibration or covered and free calibration
Excerpt from the operating menu
The following excerpt from the operating menu shows how the full and empty calibration
(for level measurements) or the covered and free calibration (for level limit detection) are
entered. The individual functions are explained in the sections below.
The functions are only accessible after the background calibration has been performed.
NOTICE
The "value full" (*14) and "value empty" (*17) functions only appear if the "level"
option was selected in the "measurement method" function (*05).
"Calibration point" (*13)
*14
value full
[%]
target value
at calibr.
point
*15
calibration
0: stop/
edit
1: start
*11
avg.
pulse
rate
*16
full calibr.
[c/s]
-1: not calibra-
ted
Group
selection
013
calibr. point
0: full/
covered
1: empty/free
*19
next point
0: no
1: yes
*17
value empty
[%]
target value
at calibr.
point
*15
calibration
0: stop/
edit
1: start
*11
avg.
pulse
rate
*18
empty calibr.
[c/s]
-1: not calibra-
ted
On-site display
Meaning
Calibr. point
*13
This function is used to select which calibration point ("full/
covered" or "empty/free") will be entered.
full/covered
empty/free
1
E
0
1
E
0