"Next point" (*19)
Options:
•
no
This option must be selected after both calibration points have been entered. After this
selection the Gammapilot M returns to the group selection and the calibration is
completed.
•
yes
This option must be selected, if only one calibration point has been entered yet. After this
selection the Gammapilot M returns to the
"calibr. point" (*13)
function and the next
point can be entered.
6.4.4
Additional settings
After the calibration, the Gammapilot M outputs the measured value via the PROFIBUS PA
interface.
Many additional functions are available for optimizing the measuring point. They can be
configured as required. A detailed description of all instrument functions is given in
Operating Instructions BA00287F/00/EN, "Gammapilot M - Description of Device
Functions". This document can be found on the supplied CD-ROM.
On-site display
Meaning
next point
*19
This function is used to specify, if a further calibration point is
to be entered or not.
no
yes