Procedure
Step
Matrix
Entry
Significance
1
V9H5
e.g. 672
Reset transmitter (default = operating mode 1)
-
Enter sensor constants at V3H5 and V3H6
2
-
»E«
Register entry
3
V0H1/V0H2 -
Calibrate as described in e.g. Section 4.2
4
V2H1
0
Select manual entry of values
5
-
»E«
Register entry
6
V2H2
1
…
30
Enter table entry number
7
-
»E«
Register entry
8
V2H3
e.g. 0
Enter volume in required units
9
-
»E«
Register entry
10
V2H4
00.00
Enter level in units used in calibration
11
-
»E«
Register entry
12
V2H5
2
…
30
Enter next table entry number
13
-
»E«
Register entry.
-
The system jumps to V2H3, V2H2 is incremented
14
V2H3
Repeat steps 8 to 13 until all points are entered
15
V2H0
3
Select manual linearization table
16
-
»E«
Press »E« to activate linearization
Note!
•
Set analogue output in units used for linearization
Manual linearization
with tabular values
Silometer FMX 570
Chapter 5: Linearization
27