Monorack wiring
Fig. 2.5 shows the layout in the base of the Monorack II housing, the pin assignments
correspond to those in Fig. 2.4. When connecting together several Monoracks, follow
the instructions supplied with the housing.
•
Set the jumper to position "Racksyst I"
•
Insert the coding pins supplied at positions 2 and 9 in the female connector
at the base of the housing.
•
The pin assignments printed in black are valid for the Silometer FMX 570.
Note!
If you are installing the Silometer FMX in a Monorack I housing, please note that there is
no jumper switch. In addition, for the 24 VDC version, the dummy card in the power
control slot must be replaced by the 24 V card supplied.
PC board
slot
BA119E14
Power
Outputs
see Fig. 2.4
Input, d2, d4
Insert coding pins
for Z-version at
Positions
2 and 9
Hole for mounting
screw
Hole for mounting
screw
Jumper positioned
for Racksyst I card
Fig. 2.5:
Layout of Monorack terminal
blocks
Silometer FMX 570
Chapter 2: Installation
12