7.2
Simulated operating mode
This function is intended primarily for checking the correct function of the system and is
selected and terminated at V8H0:
•
Enter 6 to simulate frequency, level, volume or current
•
Enter an operating mode to terminate simulation and resume normal
measurements.
Start and stop simulation
Step
Matrix
Entry
Significance
1
V8H0
e.g. 6
Selects simulation mode channel 1, entries can be made at
fields V9H6
…
V9H9
or
e.g. 0
Selects two channel measurement and ends simulation
2
-
»E«
Registers entry
Four modes are possible:
•
Simulation of frequency, V9H6
•
Simulation of level, V9H7
•
Simulation of volume, V9H8
•
Simulation of current, V9H9.
When a value is entered at the appropriate matrix, the analog outputs are fed with the
appropriate current and voltage and the other 3 simulation values are recalculated.
Throughout the simulation the red alarm LED flashes to indicate that the instrument is no
longer measuring, the alarm relay does not, however, trip.
Frequency simulation
Step
Matrix
Entry
Significance
1
V9H6
e.g. 100
Depending upon the calibration and linearization, a value
corresponding to 100 Hz is displayed
2
-
»E«
Registers entry
Level simulation
Step
Matrix
Entry
Significance
1
V9H7
e.g. 10
Depending upon the calibration and linearization, the
analogue output is fed with a current corresponding to
e.g.10 m, 10 ft, 10% level.
2
-
»E«
Registers entry
Volume simulation
Step
Matrix
Entry
Significance
1
V9H8
e.g. 100
Depending upon the calibration and linearization, the
analogue outputs are fed with the current corresponding
to 100 hl, 100 gallons, 100%
2
-
»E«
Registers entry
Current simulation
Step
Matrix
Entry
Significance
1
V9H9
e.g. 16
The analogue output is fed with a current of 16 mA and
the corresponding measured value displayed at V0H0
2
-
»E«
Registers entry
Chapter 7: Trouble-Shooting
Silometer FMX 570