Code
Type
Cause and Remedy
E 101-106 Alarm
Fault in instrument electronics
- Call Hauser Service
E 107
Alarm
Battery voltage too low
- Make back-up of entered parameters immediately
- Have battery changed at once by trained personel or ring for service
E 201-202 Alarm
Fault in probe (f < 35 Hz; f > 3000 Hz)
- Check probe and electronic insert
E 401
Alarm
Fault in probe or wiring
- Check probe, electronic insert and wiring
E 601
Warning
PFM transmission internal code check
- can be ignored if it appears only briefly
E 602
Warning
Linearization does not rise monotonously (volume does not increase
with level)
- Check and re-enter correct values, reactivate linearization
E 604
Warning
Linearization has less than two sets of values
- Enter more values, reactivate linearization
E 608
Warning
Value in V0H5 greater than that in V0H6
- Check input
E 610
Warning
Calibration fault (»empty« level > »full« level)
- Repeat calibration
E 613
Warning
Instrument in simulation mode
- Switch back when finished
Table 7.1:
Error messages
Sensor/
channel
Fault
Cause and remedy
Capacitance
Measured value wrong
•
Incorrect calibration? Check measured
value before linearisation, V0H9
- if not correct, check whether full and empty
calibration correct V0H1/V0H2
- if correct, check linearization parameters
- check operating mode, V8H0
•
Change in product
- recalibrate for new product
•
Build-up on probe
- wire electronic insert for build-up, see
Section 2.4
- clean probe
•
Probe damaged, bent or pressed to side of
vessel
- check and remedy
•
Condensation in connection compartment
Deltapilot S
Measured value wrong
•
Incorrect calibration? Check measured
value before linearisation, V0H9
- if not correct, check whether full and empty
calibration correct V0H1/V0H2
- If correct, check linearization parameters
- check operating mode, V8H0
•
Change in density of product
- recalibrate
-
•
Sensor damaged
- check and remedy
Table 7.2:
Trouble shooting table for
incorrect function without error
message
Silometer FMX 570
Chapter 7: Trouble-Shooting
34