Maintenance and care
55
7
Maintenance and care
The InnuPure C16
touch
is mainly maintenance-free. Care and maintenance are lim-
ited to cleaning the device, changing the fuses and installing software updates.
WARNING
Danger of electric shock!
The device cover must only be opened by the technical customer service of Analytik
Jena or its authorized agency. Before opening the device, disconnect the mains plug!
7.1
Cleaning the InnuPure C16 touch
WARNING
Danger of electric shock!
Prior to cleaning the device with disinfection agents, switch off the InnuPure C16
touch
and the tablet and disconnect the mains plug from the mains outlet at the rear
of the device!
The InnuPure C16
touch
must only be recommissioned once it is completely dry!
CAUTION
Risk of burns!
Wait for at least 10 minutes after switching the device off prior to start cleaning the
interior of the InnuPure C16
touch.
After this cooling-down period, the temperature
directly below the heating strip is less than 50 °C.
Avoid contamination by handling sample substances carefully.
Wipe spilled samples or reagents immediately with an absorbent cloth or piece of
paper. To do so, open the lower front door via the software or open the upper front
door manually (see Fig. 24).
If the InnuPure C16
touch
is used for the analysis of infectious material, great care
must be taken because the InnuPure C16
touch
cannot be decontaminated as a
whole device.
Visible contamination must be removed immediately using suitable detergents,
making sure that no solvent enters the inside of the device.
The adapters for accepting the reagent strips can be autoclaved and are also suita-
ble for dip and spray disinfection.
The sample tray is also suitable for dip and spray disinfection.
We recommend the following disinfection agents for the device interior
Decosept AF
by Dr Schuhmacher GmbH
Meliseptol HBV cloths
by B. Braun