InnuPure C16
touch
Safety instructions
11
2.6.5
Chemical resistance of the device
Aggressive substances may damage the device. Although the materials used are re-
sistant to most of the commonly used substances, material damage from aggressive
substances cannot be precluded.
Before using any aggressive substances (e.g. acids, bases, organic solutions), check
whether the materials in direct contact with the substances are resistant to them.
Only use substances compatible with the materials listed.
If in doubt contact Analytik Jena.
Overview 1
Components and materials in direct contact with the sample
Components
Material
Filter tips
PP
Piston seal
Aluminum, stainless steel, PTFE, rubber (NBR 70).
The pistons are not contaminated with sample if the filter tips are used.
Reagent cups
PP or PTFE
The above-mentioned components are not resistant to the following substances:
hydrogen fluoride
(HF/hydrofluoric acid)
highly concentrated acids
cleaning powder
paint thinner
naphtha
gasoline
acetone
cleaning spray
ozone
salt-containing detergents
(NaOCI/sodium hypochlorite)
2.7
Behavior during emergencies
If there is no immediate danger of injury, immediately switch the device switch of the
InnuPure C16
touch
to the "0" position in the event of dangerous situations or acci-
dents and/or disconnect the mains plug from the mains outlet!
2.8
Safety instructions: service and repair
The customer service department of Analytik Jena or its authorized and trained spe-
cialist personnel usually service the InnuPure C16
touch
.
Independent servicing can maladjust or damage the device. The operator may gener-
ally only carry out the tasks listed in chapter "Maintenance and care" p. 55 ff.