Specifications
Automatic Boat Drive
40
Electrical variables
Power supply
110 to 240 V +10/-5 %
Frequency
50/60 Hz
Overvoltage category
II
Degree of contamination
2
Fuses
2 x T 3.15 A H
Max. power consumption
55 VA
Analyzer interface
RS 232 (special bus)
Environmental conditions
Temperature during stor-
age
+15 to +55 °C
Temperature during oper-
ation
+20 to +35 °C
Humidity during operation
max. 90 % at 30 °C
Humidity during storage
10 to 30 % (use desiccant)
Air pressure
0.7 to 1.06 bar
Maximum altitude
2000 m
10.1 Standards and directives
Protection class and protection
type
The device is protection class I. The housing is protection type IP 20.
Device safety
The device complies with the following safety standards
¡
EN 61010-1
¡
EN 61010-2-081
¡
EN 61010-2-010
EMC compatibility
The device has been checked for transient emissions and noise immunity.
It meets the requirements for transient emissions according to
¡
EN 61326-1 (EN 55011 group 1, class B)
The device meets the requirements for noise immunity according to
¡
EN 61326-1 (requirements for use in a basic environment)
Environmental compatibility
The device has been tested for environmental compatibility and meets the requirements
according to
¡
ISO 9022-2
¡
ISO 9022-3
EU directives
The device meets the requirements of the directive 2011/65/EU.
The device is designed and tested in accordance with standards meeting the require-
ments of EU directives 2014/35/EU and 2014/30/EU. The device leaves the factory in a
sound condition with regard to technical safety. To maintain this condition and to en-
sure safe operation, the user must strictly observe the safety and operating instructions
contained in this operating manual. For accessories delivered with the device and sys-
tem components from other manufacturers, the information provided in their respective
operating manuals has priority.