Transport and storage
Automatic Boat Drive
34
8.1.3
Moving the ABD in the laboratory
NOTICE
Falling out of loose parts as well as unintentional dropping of the ABD creates a
risk of injury and damage to the device.
Before moving the ABD, remove all loose parts and dismount the sampler from the ABD
if necessary!
Move the ABD with great care! Securely hold the ABD with both hands from below.
To move the ABD within the laboratory:
}
Switch off the ABD from the equipment switch and disconnect the mains plug from
the mains outlet.
}
Remove the interface cable ("RS 232 multi EA" connector) and the flame sensor cable
("flame" connector).
}
Disconnect the gas supply and remove the connection hose 16 from the ABD.
}
If necessary, remove the sampler from the ABD and move it separately in the lab (see
user manual of the sampler).
}
Disconnect the ABD from the furnace of the basic module (→ "Disconnecting the
ABD from the combustion tube"
27).
Securely hold the device with both hands from below.
For installation at the new location, observe the location requirements (→ "Location re-
quirements"
8.2
Storage
NOTICE
Environmental influences and condensate formation can destroy individual
components of the ABD!
The ABD may only be stored in air conditioned rooms. The atmosphere must be low in
dust and free from aggressive vapors.
If the ABD is not installed immediately after delivery or not needed for prolonged peri-
ods, it should be stored in its original packaging. A suitable desiccant should be added to
the packaging to prevent damage from moisture.
The following requirements are placed on the climatic conditions in the storage room of
the ABD:
¡
Temperature range: +15 to +55 °C
¡
Max. humidity: 10 to 30 % (use desiccant)
¡
Air pressure: 0.7 to 1.06 bar