6.2.3
Analytic problems
Error
Possible cause
Remedy
Carryover
Sample sluice contami-
nated with sample
Remove sample material
from the sample sluice,
wipe out the sluice, lid
and injection port with al-
cohol and allow them to
dry.
ABD connection piece
contaminated
Clean the ABD connection
piece and seal elements
and replace them if nec-
essary.
Inadequate cooling of the
sample sluice
Check cooling, if neces-
sary inform Service