Maintenance and care
Automatic Boat Drive
28
NOTICE
Pay special attention to the connection cables on the rear of the ABD (for flame sensor
and sampler cable). These could easily tear off.
}
Only carry out the following if necessary during maintenance:
Loosen the clamping ring at the sample sluice.
}
Remove the ABD connection piece from the sample sluice.
7.4
Replacing the septum and cleaning the sluice at the ABD
CAUTION
Crushing risk when closing the sample sluice!
High forces occur when closing the sample sluice of the ABD. Do not reach into the
sluice area during closing.
Replace septum
If necessary, replace the septum at the injection port of the sample sluice as follows:
1
3
2
}
Unscrew the cap (2) from the injection port (1) and remove the used
septum (3) from the cap.
}
Insert a new septum into the cap.
}
Screw the cap onto the injection port and tighten it finger-tight.
ü
The septum has been replaced.
Clean sluice
}
Open the sample sluice. Push the "port" button at the ABD.
}
Clean the interior of the sluice with cellulose soaked in Ethanol.
Then allow the sluice to dry out well.
}
Close the sluice by pushing the "port" button again.
}
The sluice closes.