4. Only the Emerson Process Management 375 or 475 Field Communicator is
approved for use with this transmitter.
5. The maximum permitted operating temperature of the Rosemount
™
3308A
transmitter is 70°C. To avoid the effects of process temperature and other thermal
effects care shall be taken to ensure that the “Electronics Temperature” does not
70°C.
B.7
Canada
B.7.1
I6 Canada Intrinsically Safe
Certificate
FM17CA0007X
Standards
C22.2 No. 94-M91:1991 (R2011), CAN/CSA C22.2 No. 60079-0:2015,
CAN/CSA 22.2 60079-11:2014, C22.2 No 61010-1:2004, C22.2 No.
60529:2016
Markings
Intrinsically Safe
Class I, Division 1, Groups A, B, C and D T4; (-55°C ≤ Ta ≤ +70°C)
Ex ia IIC Ga T4; (-55°C ≤ Ta ≤ +70°C)
when installed per Rosemount drawing 03308-1010;
Type 4X; IP66/67
Special Conditions for Safe Use (X):
1. The Model 3308 transmitter housing contains aluminum; protect the enclosure to
avoid a potential risk of ignition due to impact or friction.
2. The surface resistivity of the polymeric antenna is greater than 1GΩ. To avoid
electrostatic charge build-up, it shall not be rubbed or cleaned with solvents or a dry
cloth.
3. For use with only the Emerson Process Management Model 701PBKKF SmartPower
™
Option or the Computational Systems, Inc Model MHM-89004 battery module.
4. Only the Emerson Process Management 375 or 475 Field Communicator is
approved for use with this transmitter.
5. The maximum permitted operating temperature of the Rosemount
™
3308A
transmitter is 70°C. To avoid the effects of process temperature and other thermal
effects care shall be taken to ensure that the “Electronics Temperature” does not
70°C.
B.8
Europe
B.8.1
I1 ATEX Intrinsic Safety
Certificate
FM 12ATEX0072X
Reference Manual
Product Certifications
00809-0100-4308
February 2020
Reference Manual
189
Summary of Contents for Rosemount 3308A
Page 8: ...Contents Reference Manual February 2020 00809 0100 4308 viii Rosemount 3308 Series ...
Page 10: ...Introduction Reference Manual February 2020 00809 0100 4308 10 Rosemount 3308 Series ...
Page 22: ...Transmitter overview Reference Manual February 2020 00809 0100 4308 22 Rosemount 3308 Series ...
Page 74: ...Installation Reference Manual February 2020 00809 0100 4308 74 Rosemount 3308 Series ...
Page 96: ...Configuration Reference Manual February 2020 00809 0100 4308 96 Rosemount 3308 Series ...