![Emerson Rosemount 326P Quick Start Manual Download Page 41](http://html.mh-extra.com/html/emerson/rosemount-326p/rosemount-326p_quick-start-manual_104710041.webp)
5
Troubleshooting
The pressure transmitter monitors itself for issues then reports the issues using the
display. If the pressure transmitter detects a fault, it will react according to FOU1 and
FOU2. Use the information in this section to identify and resolve common issues with the
pressure transmitter.
5.1
Error messages
Use the information in the table below to resolve errors with the pressure transmitter.
Display
Description
Corrective actions
Off
The supply voltage is too low (fault
number W403).
•
Check and/or correct the supply
voltage.
•
In a two-wire operation, check
and/or correct the connected
load.
PARA
The parameter setting is outside the
permitted range.
Reset the parameter.
E100
The pressure transmitter detected an
internal error.
Replace the pressure transmitter.
5.2
Warning messages
Use the information in the table below to resolve issues indicated by warning messages
with the pressure transmitter.
Display
LED Status
Description
Corrective actions
SC1
OUT1 Flashes
Excessive current switching
output 1.
Check switching output 1 for a
short circuit or excessive
current then remove the fault.
SC2
OUT2 Flashes
Excessive current switching
output 2.
Check switching output 2 for a
short circuit or excessive
current then remove the fault.
SC
OUT1 and
OUT2 are
flashing
Excessive current switching
output 1 and switching output
2.
Check switching output 1 and 2
for a short circuit or excessive
current then remove the fault.
OL
The measured value is more
than five percent above the
upper range limit (URL).
Check the system pressure and
correct it if necessary.
UL
The measured value is more
than five percent below the
lower range limit (LRL).
Check the system pressure and
correct it if necessary.
Quick Start Guide
Troubleshooting
00825-0100-4327
January 2019
Quick Start Guide
41
Summary of Contents for Rosemount 326P
Page 1: ...Quick Start Guide 00825 0100 4327 Rev 01 January 2019 Rosemount 326P Pressure Transmitter ...
Page 2: ...2 ...
Page 4: ...Contents Quick Start Guide January 2019 00825 0100 4327 4 Quick Start Guide ...
Page 14: ...Installation Quick Start Guide January 2019 00825 0100 4327 14 Quick Start Guide ...
Page 40: ...Setting ranges Quick Start Guide January 2019 00825 0100 4327 40 Quick Start Guide ...
Page 44: ...Product certifications Quick Start Guide January 2019 00825 0100 4327 44 Quick Start Guide ...
Page 47: ...Quick Start Guide 00825 0100 4327 January 2019 Quick Start Guide 47 ...