The extended functions menu opens.
3. Select
SIM
from the menu.
4. Press
Set
.
The simulation menu opens. The first option is
SEL
.
Set the simulation output state
Follow these steps to set the output state for a simulation.
Procedure
1. Select
SEL
from the simulation menu.
2. Press and hold
Set
to change the simulation menu to
OU
.
3. For three-wire operation, select
S.OU1
from the simulation menu.
For two-wire operation, skip to step
4. Press and hold
Set
to set the value.
•
OPEN
: output 1 not active/open
•
CLOS
: output 1 active/closed
5. Press
Mode/Enter
.
6. Select
S.OU2
from the simulation menu.
7. Press and hold
Set
to set the value.
•
If
OU2
=
Hnc
,
Hno
,
Fnc
or
Fno
(not in two-wire operation):
—
OPEN
: output 2 not active/open.
—
CLOS
: output 2 active/closed.
•
If
OU2
=
I
or
InEG
:
—
3.60-21.10 mA in steps of 0.01 mA.
8. Press
Mode/Enter
.
Set the simulation pressure value
The pressure transmitter will display the simulation pressure value during the simulation.
Follow these steps to set the pressure value for the simulation.
Procedure
1. Select
SEL
from the simulation menu.
2. Press and hold
Set
to change the simulation menu to
Proc
.
3. Press
Mode/Enter
.
The pressure transmitter displays
S.Pr
.
4. Select
S.Pr
from the simulation menu.
5. Press and hold
Set
to modify the pressure value.
6. Press
Mode/Enter
.
Quick Start Guide
Operation
00825-0100-4327
January 2019
Quick Start Guide
31
Summary of Contents for Rosemount 326P
Page 1: ...Quick Start Guide 00825 0100 4327 Rev 01 January 2019 Rosemount 326P Pressure Transmitter ...
Page 2: ...2 ...
Page 4: ...Contents Quick Start Guide January 2019 00825 0100 4327 4 Quick Start Guide ...
Page 14: ...Installation Quick Start Guide January 2019 00825 0100 4327 14 Quick Start Guide ...
Page 40: ...Setting ranges Quick Start Guide January 2019 00825 0100 4327 40 Quick Start Guide ...
Page 44: ...Product certifications Quick Start Guide January 2019 00825 0100 4327 44 Quick Start Guide ...
Page 47: ...Quick Start Guide 00825 0100 4327 January 2019 Quick Start Guide 47 ...