Table 4-2: rP1 and rP2
(continued)
Unit
Minimum
Maximum
Step increment
Pressure range 2A
inH
2
O
-5.0
99.8
0.1
mbar
-12.4
248.4
0.2
kPa
-1.24
24.84
0.02
mmWS
-126
2536
2
Pressure range 3A
psi
-14.50
14.45
0.05
inH
2
O
-401
400
1
mbar
-1000
998
1
kPa
-100.0
99.8
0.1
mWS
-10.20
10.18
0.01
Pressure range 2B
psi
-0.73
14.47
0.01
inH
2
O
-20.0
400.8
0.4
mbar
-50
998
1
kPa
-5.0
99.8
0.1
mWS
-0.51
10.18
0.01
Pressure range 4A
psi
-1.44
19.96
0.02
inH
2
O
-40
636
0.5
mbar
-100
1582
2
kPa
-10.0
158.2
0.2
mWS
-1.02
16.06
0.02
Pressure range 5A
psi
-1.80
35.94
0.03
inH
2
O
-50
995
1
bar
-0.124
2.478
0.002
kPa
-12.4
247.8
0.2
mWS
-1.26
25.29
0.01
Pressure range 6A
psi
-14.50
54.9
0.05
bar
-1.000
3.78
0.005
kPa
-100.0
378
0.5
Quick Start Guide
Setting ranges
00825-0100-4327
January 2019
Quick Start Guide
35
Summary of Contents for Rosemount 326P
Page 1: ...Quick Start Guide 00825 0100 4327 Rev 01 January 2019 Rosemount 326P Pressure Transmitter ...
Page 2: ...2 ...
Page 4: ...Contents Quick Start Guide January 2019 00825 0100 4327 4 Quick Start Guide ...
Page 14: ...Installation Quick Start Guide January 2019 00825 0100 4327 14 Quick Start Guide ...
Page 40: ...Setting ranges Quick Start Guide January 2019 00825 0100 4327 40 Quick Start Guide ...
Page 44: ...Product certifications Quick Start Guide January 2019 00825 0100 4327 44 Quick Start Guide ...
Page 47: ...Quick Start Guide 00825 0100 4327 January 2019 Quick Start Guide 47 ...