26
Manually modifying the Calibration Parameters
Section 3: Meter maintenance
Reference, Installation and Operation Manual
December 2017
3-9000-764 Rev F
Chord “L” Dimension Calculation
The chord “L” dimension is calculated from the meter housing length and the transducer
housing lengths as shown in
. The lengths of t he transducer housings are etched on
on the individual components. The length of the meter body is found on the original calibration
sheet supplied with the meter.
Equation 3-2 Chord “L” Dimension
where
=
chord “L” dimension (in) (
LA
...
LD
)
=
meter housing length (in)
=
transducer 1 housing length (in)
=
transducer 2 housing length (in)
Lchord
L
MeterHou
g
sin
L
Hsng
1
L
Hsng
2
–
–
=
L
chord
L
MeterHou
g
sin
L
Hsng
1
L
Hsng
2