Maintenance and Troubleshooting Manual
Section 2: Troubleshooting
3-9000-764 Rev F
December 2017
Meter maintenance
23
Meter Monitor (Detailed) view
Meter Data List
•
Electronics Temperature out of
range
•
Temperature of the electronics is out
of nominal operating range
below -40 °C or above 100 °C (-40 °F
or above 212 °F).
—
Heat or cool the meter electron-
ics housing. If operating
temperature exceeds 65
o
C,
remote mount the Transmitter
Electronics Enclosure.
Meter Monitor (Detailed) view
Meter Data List
•
Electronics voltage out of range
•
Systems voltages are valid if 1.0V,
1.2V, 2.5V, 3.3V or the Acquisition
Module valid voltages are 1.2V, 2.5V
or 3.3V.
•
Replace the CPU Module if one or
more of the System Voltages is out of
range.
•
Replace the Acquisition Module if one
or more of the voltages is out of
range.
MeterLink Tools Menu‘
•
Frequency output
•
Run the Frequency Outputs test
•
If the output reads zero, you may
require a pull up resistor 1.2kOHM,
0.5W.
Table 2-2 Maintenance
Daniel MeterLink utility
Diagnostics
Action(s)