Maintenance and Troubleshooting Manual
Section 2: Troubleshooting
3-9000-764 Rev F
December 2017
Meter maintenance
25
Meter Hardware
Diagnostics
Action(s)
Meter Electronics
•
Acquisition Module communications error
•
Check firmware revision and upgrade if
necessary using Daniel MeterLink Tools>Pro-
gram Download.
•
If the CPU Module LED 5 is not flashing green,
check interconnect cable between acquisition
Module and the CPU Module.
•
If the CPU Board LED 5 is not flashing green,
replace Acquisition Module.
MeterLink Logs/
Reports Menu
•
Meter performed a Warm Start or a Warm
start required
•
Meter performed a Warm Start:
—
Collect an Archive event log (Audit log)
using Daniel MeterLink to view configu-
ration parameters changes and when
they changed.
•
Warm start required:
—
When you make changes to the trans-
ducer characteristics, sample rates, the
device number, or a Modbus map file.