26
Meter maintenance
Section 2: Troubleshooting
Maintenance and Troubleshooting Manual
December 2017
3-9000-764 Rev F
Meter Hardware
Diagnostics
Action(s)
MeterLink Tools>Edit/
Compare Configura-
tion Menu
•
Meter performed a Cold Start
•
The meter configuration has reset to default
values and the meter is not configured
correctly to measure flow.
•
Unless the cold start occurred after upgrading
firmware, you may need to replace the CPU
Module.
•
If the Cold Start occurred after a firmware
upgrade, you must reconfigure the meter
from a previously saved configuration file
using the Edit>Compare Configuration screen.
Then clear the latched alarm on the Status
Summary page.
MeterLink Logs/
Reports Menu
•
Power failure
•
If this was a known power fail or restart of the
meter just acknowledge this alarm on the
Status Summary page.
•
If this was an unexpected restart of the meter,
verify the integrity of the power to the meter
and make sure that the voltage level is in the
range of 11-36 VDC at the meter.
Meter Hardware
Diagnostics
Action(s)
Daniel MeterLink
Meter Monitor
(Summary) view
•
Chord Failure Daniel MeterLink
Meter Monitor (Summary) view
•
The meter is unable to obtain measurement
data from a pair of transducers.
•
The cause may be isolated to one pair of trans-
ducers or its cabling. Check the transducer
wiring for this pair of transducers to make sure
connections are secure and wired correctly.
•
Verify that the meter run is not partially full
where this top transducer pair is not
submerged in the process fluid.
•
Verify the average gain of this transducer pair
is not above 90dB. Read the value from the
Daniel MeterLink Monitor Page or using AMS
under Service Tools|Path performance.
•
Remove the transducer and clean the trans-
ducer face. Reapply coupling fluid to the
transducer face and reinstall.(
Transducer removal procedure).