SECTION 5.0
PROGRAMMING THE ANALYZER
5.1 GENERAL
This section describes how to do the following:
1.
configure and assign values to the current outputs
2.
configure and assign setpoints to the alarm relays
3.
choose turbidity or total suspended solids (TSS)
4.
set a security code
5.
tell the analyzer the frequency of the ac power (needed for optimum noise rejection)
6.
tell the analyzer the number of sensors being used
7.
reset the analyzer to factory calibration and default settings
8.
select a default display screen
Default settings are shown in Table 5-1 on the following page. To change a default setting, refer to the section list-
ed in the table. To reset default settings, see Section 5.9.
5.2 CHANGING STARTUP SETTINGS
When the Solu Comp II is powered up for the first time, startup screens appear. The screens prompt the user to
identify the number of sensors being used, the measurement (turbidity or TSS), and the units. If incorrect settings
were entered at startup, enter the correct settings now. To change the number of sensors refer to Section 5.6. To
change the measurement, refer to Section 5.5.
FOR BEST RESULTS, ENTER THE NUMBER OF SENSORS BEING USED
(SECTION 5.6) AND WHETHER TURBIDITY OR TSS IS BEING MEASURED
(SECTION 5.5) BEFORE MAKING OTHER PROGRAM SETTINGS.
MODEL CLARITY II
SECTION 5.0
PROGRAMMING THE ANALYZER
19